The effect of implant number and position on the stress behavior of mandibular implant retained overdentures: A three-dimensional finite element analysis

dc.authorid0000-0001-6650-8067en_US
dc.contributor.authorTopkaya, Tolga
dc.contributor.authorSolmaz, Murat Yavuz
dc.date.accessioned2021-11-24T08:31:51Z
dc.date.available2021-11-24T08:31:51Z
dc.date.issued2015-07-16en_US
dc.departmentBatman Üniversitesi Mühendislik - Mimarlık Fakültesi Makine Mühendisliği Bölümüen_US
dc.description.abstractThe present study evaluated the effects of ball anchor abutment attached to implants with a 4.30 mm diameter and 11 mm insert length on stress distribution in a patient without any remaining teeth in the lower jaw. In the study, the stress analysis was performed for five different configurations (2 with 4 implant-supported and 3 with 2 implant-supported) and three different loading types using ANSYS Workbench software. The stresses measured in the 4 implant-supported models were lower compared to the stresses measured in the 2 implant-supported models. The stresses on the implants intensified on the cervical region of the implants. When the effects of the loading sites on the stress were examined, the loading on the first molar tooth produced the highest stresses on the implants.en_US
dc.identifier.citationTopkaya, T., Solmaz, M.Y. (2015). The effect of implant number and position on the stress behavior of mandibular implant retained overdentures: A three-dimensional finite element analysis. Journal of Biomechanics, 48 (10), pp.2102-2019. DOI:https://dx.doi.org/10.1016/j.jbiomech.2015.03.006en_US
dc.identifier.endpage2109en_US
dc.identifier.issn00219290
dc.identifier.issue10en_US
dc.identifier.scopusqualityN/Aen_US
dc.identifier.startpage2102en_US
dc.identifier.urihttps://dx.doi.org/10.1016/j.jbiomech.2015.03.006
dc.identifier.urihttps://hdl.handle.net/20.500.12402/4060
dc.identifier.volume48en_US
dc.identifier.wosqualityN/Aen_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.language.isoenen_US
dc.publisherElsevieren_US
dc.relation.isversionof10.1016/j.jbiomech.2015.03.006en_US
dc.relation.journalJournal of Biomechanicsen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/restrictedAccessen_US
dc.rightsAttribution-NonCommercial-ShareAlike 3.0 United States*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-sa/3.0/us/*
dc.subjectDental Implanten_US
dc.subjectRemovable Overdentureen_US
dc.subjectFinite Element Method (FEM)en_US
dc.titleThe effect of implant number and position on the stress behavior of mandibular implant retained overdentures: A three-dimensional finite element analysisen_US
dc.title.alternativeMandibular implant destekli diş üstü protezlerin stres davranışına implant sayısı ve pozisyonunun etkisi: Üç boyutlu bir sonlu eleman analizien_US
dc.typeConference Objecten_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
Ä°sim:
The effect of implant number and position on the stress behavior of mandibular implant retained overdentures.pdf
Boyut:
8.47 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text
Lisans paketi
Listeleniyor 1 - 1 / 1
Küçük Resim Yok
Ä°sim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: